Measurement of total lung capacity
plethysmography versus single breath helium method after treatment with bronchodilator
DOI:
https://doi.org/10.22491/2357-9730.125264Keywords:
Pulmonary function tests, lung volume, airflow obstructive defectAbstract
OBJECTIVE: Total lung capacity (TLC) can be measured using body plethysmography,
single-breath helium or RX planimetry. Body plethysmography is the goldstandard test,
allowing comparison with other methods. The single breath helium method (SBHe),
when used in association with CO2
diffusing capacity, can estimate lung volumes in
patients with airway flow obstruction. The aim of the present study is to evaluate if TLC
measured with SBHe after treatment with bronchodilator is similar to that obtained from
body plethysmography.
MATERIALS AND METHODS: The sample included 155 adult patients submitted to
pulmonary function testing including lung volumes at the Pulmonary Physiology Unit,
Pulmonary Department at Hospital de Clínicas de Porto Alegre. The patients were placed
in two groups: one group underwent SBHe with bronchodilator, and the other group was
submitted to SBHe without bronchodilator. Afterwards, each group was subdivided into
two subgroups: normal patients and patients with airway obstruction.
RESULTS: The TLC results obtained with the body plethysmography technique were
greater than the alternative methods for all groups. In the group without bronchodilator
treatment, there was a median difference of 934 ml (CI 95% = 657 ml – 1210 ml;
P<0.001) for normal patients, and a median difference of 1679 ml (95% Cl = 1289ml –
2069 ml; P<0.001) for patients with airway obstruction. While bronchodilator treatment
was used, the median difference for normal patients with airway obstruction was 1751
ml (95% Cl 1516 – 1985 ml; P<0.001).
CONCLUSIONS: The results showed that treatment with bronchodilator did not make
TLC measurements similar when SBHe was compared to body plethysmography, neither
for normal patients nor for patients with airway obstruction; this leads to the conclusion
that SBHe is not a reliable test for TLC measurements.
Downloads
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution 4.0 International License.
Authors who publish with this journal agree to the following terms:
Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).