This is an outdated version published on 19-12-2025. Read the most recent version.

Analytical Method Validation for the Detection of the Drug Sirolimus in Sugar-Coated Tablet using HPLC-DAD

Authors

DOI:

https://doi.org/10.22456/2527-2616.150632

Keywords:

sirolimus, quality control, liquid chromatography

Abstract

Sirolimus (SRL) is an immunosuppressive macrolide widely used in transplant patients to prevent organ rejection. The development of an analytical method with adequate selectivity, sensitivity, and reliability is essential for its quality control and therapeutic monitoring. A validated analytical method for the quantification of SRL was developed using High-Performance Liquid Chromatography coupled with a Diode Array Detector (HPLC-DAD). The method demonstrated linearity for SRL in the concentration range of 10 to 80 μg.mL⁻¹ (r = 0.9999). Precision was confirmed with coefficients of variation below 2%, and accuracy showed recovery values close to 100%. The validated method was successfully applied to sugar-coated tablets, demonstrating its suitability for routine quality control. Therefore, the developed method can be reliably used for the quantitative analysis of Sirolimus in sugar-coated tablets.

Downloads

Download data is not yet available.

References

Moes DJAR, Guchelaar HJ, de Fijter JW. Sirolimus and everolimus in kidney transplantation. Drug Discovery Today. 2015;20(10):1243-1249. doi:10.1016/j.drudis.2015.05.006. DOI: https://doi.org/10.1016/j.drudis.2015.05.006

Holt DW, Lee T, Johnston A. Measurement of sirolimus in whole blood using high-performance liquid chromatography with ultraviolet detection. Clin Ther. 2000 Jan;22:B38–48. DOI: https://doi.org/10.1016/S0149-2918(00)89021-9

Jones K, Saadat-Lajevardi S, Lee T, Horwatt R, Hicks D, Johnston A, et al. An immunoassay for the measurement of sirolimus. Clin Ther. 2000 Jan;22:B49–61. DOI: https://doi.org/10.1016/S0149-2918(00)89022-0

Pieri M, Miraglia N, Castiglia L, Genovese G, Basilicata P, Simonelli A, et al. Determination of rapamycin: quantification of the sodiated species by an ion trap mass spectrometer as an alternative to the ammoniated complex analysis by triple quadrupole. Rapid Communications in Mass Spectrometry. 2005 Nov 15;19(21):3042–50. DOI: https://doi.org/10.1002/rcm.2168

Rao MVB, Nagendrakumar AVD. Development and Validation of RP-HPLC Method for the Estimation of Sirolimus in Formulation and Serum [Internet]. IDC Online; [date unknown] [cited 2025 Sep 29]. Available from:https://idc-online.com/technical_references/pdfs/chemical_engineering/DEVELOPMENT%20AND%20VALIDATION.pdf

Brazil. National Health Surveillance Agency (ANVISA). Resolution RDC No. 166, of July 24, 2017. Provides for the validation of analytical methods and other provisions. Official Gazette of the Federal Government of Brazil. July 25, 2017; Sec. 1:87.

Il’ichev YV, Alquier L, Maryanoff CA. Degradation of rapamycin and its ring-opened isomer: role of base catalysis. Arkivoc. 2007;2007(12):110–31. DOI: https://doi.org/10.3998/ark.5550190.0008.c09

Asrani SK, Leise MD, West CP, Murad HM, Pedersen RA, Erwin PJ, et al. Use of sirolimus in liver transplant recipients with renal insufficiency: a systematic review and meta-analysis. Hepatology. 2010;52(4):1360–70. DOI: https://doi.org/10.1002/hep.23835

Augustine JJ, Bodziak KA, Hricik DE. Use of sirolimus in solid organ transplantation. Drugs. 2007;67(3):369–91. DOI: https://doi.org/10.2165/00003495-200767030-00004

Keogh A, Richardson M, Ruygrok P, Spratt P, Galbraith A, O’Driscoll G, et al. Sirolimus in de novo heart transplant recipients reduces acute rejection and prevents coronary artery disease at 2 years. Circulation. 2004;110(17):2694–700 DOI: https://doi.org/10.1161/01.CIR.0000136812.90177.94

Yanik EL, Siddiqui K, Engels EA. Sirolimus effects on cancer incidence after kidney transplantation: a meta-analysis. Cancer Med. 2015;4(9):1448–59 DOI: https://doi.org/10.1002/cam4.487

Islambulchilar Z, Ghanbarzadeh S, Emami S, Valizadeh H, Zakeri-Milani P. Development and validation of an HPLC method for the analysis of sirolimus in drug products. Adv Pharm Bull. 2012;2(2):135–9.

Campos MST, Fialho SL, Pereira BG, Yoshida MI, Oliveira MA. Kinetics studies of the degradation of sirolimus in solid state and in liquid medium. J Therm Anal Calorim. 2017;130(3):1653–61. DOI: https://doi.org/10.1007/s10973-017-6580-1

Peris-Vicente J, Esteve-Romero J, Carda-Broch S. Validation of analytical methods based on chromatographic techniques: an overview. In: Poole CF, editor. Analytical separation science. Weinheim (Germany): Wiley; 2015. p. 1757–808. DOI: https://doi.org/10.1002/9783527678129.assep064

Downloads

Published

19-12-2025

Versions

How to Cite

Almeida Oschelski, N., Venturi, C. R., Oliveira de Almeida, S. H., Ayres, M. V., & Steppe, M. (2025). Analytical Method Validation for the Detection of the Drug Sirolimus in Sugar-Coated Tablet using HPLC-DAD. Drug Analytical Research, 9(2), 57–62. https://doi.org/10.22456/2527-2616.150632

Issue

Section

ORIGINAL ARTICLES