Blood Glucose Measurement in Rabbits - Performance Evaluation of Portable Glucometers
DOI:
https://doi.org/10.22456/1679-9216.140309Keywords:
glycemia, glycemic range, biochemical analyzer, devices, Oryctolagus cuniculus, lagomorphs, PGMsAbstract
Background: Blood glucose level measurement plays an important role in the diagnosis and prognosis of different clinical conditions in rabbits. Portable glucometers offer a quick and affordable tool to measure blood glucose. However, not all commercially available glucometers are suitable for this purpose in different animal species and in different clinical conditions. This study aimed to assess the performance of 4 portable glucometers compared to a biochemical analyzer using blood samples from both healthy and sick rabbits (Oryctolagus cuniculus).
Material,s Methods & Results: Blood samples were collected via jugular venipuncture from 51 rabbits. The blood glucose concentration was measured using 3 human (FO= FreeStyle Optium Neo®; AA= Accu-chek® Active; IS= Injex Sens II®) and 1 veterinarian (PGM= Wellion Vet®, Gluco Calea) glucometer, configured for dog (WD), cat (WC), and horse (WH). The results obtained from the glucometers were compared to those from the laboratory biochemical analyzer in wet biochemistry, using Spearman correlation, intraclass correlation coefficient, Bland-Altman analysis, error grid analysis and total error analysis. Blood glucose concentrations were underestimated in the human devices evaluated and overestimated in the veterinary device. Bland-Altman analysis revealed significant systematic differences in 3 of the 4 devices tested. Total error above 20% was observed in normoglycemic samples from 2 devices (FO; and PGM - WC= cat code) and in hyperglycemic samples from 1 glucometer (AA). All glucometers showed glycemic results within Zone A and B of the error grid analysis, although a glucometer (FO) present most of the results (60.8%) in Zone B. Only a single portable glucometer (IS) consistently met the objectives across all analyses.
Discussion: The sample included 26 males and 25 females, with ages ranging from 5 to 96 months. Thirty-five rabbits were healthy, while 16 had various clinical abnormalities. Significant differences in median glycemia were found between the biochemical analyzer and all PGMs, except for the IS glucometer. FO and AA showed significantly lower glycemic values, while WD, WC, and WH showed higher values. The IS glucometer demonstrated the smallest mean variation and highest intraclass correlation coefficient (ICC), indicating excellent agreement. Bland-Altman plots revealed that FO and AA underestimated glycemic values, while WD, WC, and WH overestimated them. Only IS showed agreement with the laboratory analyzer. Error grid analysis indicated no serious errors with any PGMs; however, IS had over 75% of results in Zone A, showing high agreement, while FO had the least favorable results. Total error analysis suggested that only IS consistently met accuracy objectives across various analyses. The study concluded that IS was the most accurate PGM, while others showed varying degrees of inaccuracy. Although PGMs are clinically acceptable based on ISO 15197:2013 standards, accuracy and precision issues remain, particularly for euglycemic and hyperglycemic samples. Although all glucometers showed clinically acceptable results in the error grid analysis, the accuracy varied among the devices. Collectively, the IS glucometer achieved the best results in measuring blood glucose in pet rabbits, demonstrating a better accuracy and lower tendency for systematic error compared to the laboratory analyzer. Despite the lower accuracy, the other evaluated glucometers showed potential for use in rabbits when a laboratory analyzer is not available, but should be used with caution, considering the glycemic range presented.
Keywords: glycemia, glycemic range, biochemical analyzer, Oryctolagus cuniculus, lagomorphs, diagnosis.
Downloads
References
Bluwol K., Duarte R., Lustoza M.D., Simões D.M.N. & Kogika M.M. 2007. Avaliação de dois sensores portáteis para mensuração da glicemia em cães. Arquivos Brasileiros de Medicina Veterinária e Zootecnia. 59(6): 1408-1411. DOI: 10.1590/S0102-09352007000600009. DOI: https://doi.org/10.1590/S0102-09352007000600009
Brito-Casillas Y., Figueirinhas P., Wiebe J.C., López-Ríos L., Pérez-Barreto D., Argüello A., Flores C., Batista M., Camacho M., Martín A., Rodríguez J.M. & Melián C. 2014. ISO-Based assessment of accuracy and precision of glucose meters in dogs. Journal of Veterinary Internal Medicine. 28(5): 1405-1413. DOI: 10.1111%2Fjvim.12397. DOI: https://doi.org/10.1111/jvim.12397
Cohen T.A., Nelson R.W., Kass P., Christopher M.M. & Feldman E.C. 2009. Evaluation of six portable blood glucose meters for measuring blood glucose concentrations in dogs. Journal of the American Veterinary Medical Association. 235(3): 276-280. DOI: 10.2460/javma.235.3.276. DOI: https://doi.org/10.2460/javma.235.3.276
Cohn L.A., McCaw D.L., Tate D.J. & Johnson J.C. 2000. Assessment of five portable blood glucose meters, a point-of-care analyzer, and color test strips for measuring blood glucose concentration in dogs. Journal of the American Veterinary Medical Association. 216(2): 198-202. DOI: 10.2460/javma.2000.216.198. DOI: https://doi.org/10.2460/javma.2000.216.198
Domori A., Sunahara A., Tateno M., Miyama T.S., Setoguchi A. & Endo Y. 2014. The clinical utility of two human portable blood glucose meters in canine and feline practice. Veterinary Clinical Pathology. 43(1): 55-62. DOI: 10.1111/vcp.12115. DOI: https://doi.org/10.1111/vcp.12115
Gerber K.L. & Freeman K.P. 2016. ASVCP guidelines: quality assurance for portable blood glucose meter (glucometer) use in veterinary medicine. Veterinary Clinical Pathology. 45(1): 10-27. DOI: 10.1111/vcp.12310. DOI: https://doi.org/10.1111/vcp.12310
Hackett E.S. & McCue P.M. 2010. Evaluation of a veterinary glucometer for use in horses. Journal of Veterinary Internal Medicine. 24(3): 617-621. DOI: 10.1111/j.1939-1676.2010.0481.x. DOI: https://doi.org/10.1111/j.1939-1676.2010.0481.x
Harcourt-Brown F.M. & Harcourt-Brown S.F. 2012. Clinical value of blood glucose measurement in pet rabbits. Veterinary Record. 170(26): 674. DOI: 10.1136/vr.100321. DOI: https://doi.org/10.1136/vr.100321
Harr K.E., Flatland B., Nabity M., Freeman K.P. & ASVCP. 2013. ASVCP guidelines: allowable total error guidelines for biochemistry. Veterinary Clinical Pathology. 42(4): 424-436. DOI: 10.1111/vcp.12101. DOI: https://doi.org/10.1111/vcp.12101
Hollis A.R., Dallap Schaer B.L., Boston R.C. & Wilkins P.A. 2008. Comparison of the Accu-Chek Aviva point-of-care glucometer with blood gas and laboratory methods of analysis of glucose measurement in equine emergency patients. Journal of Veterinary Internal Medicine. 22(5): 1189-1195. DOI: 10.1111/j.1939-1676.2008.0148.x. DOI: https://doi.org/10.1111/j.1939-1676.2008.0148.x
International Organization for Standardization. 2013. In vitro diagnostic test systems - requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. European committee for standardization (CEN), ISO, Geneva, Switzerland, DIN EN ISO 15197: 2013.
Kang M.H., Kim D.H., Jeong I.S., Choi G.C. & Park H.M. 2016. Evaluation of four portable blood glucose meters in diabetic and non-diabetic dogs and cats. Veterinary Quarterly. 36(1): 2-9. DOI: 10.1080/01652176.2015.1092617. DOI: https://doi.org/10.1080/01652176.2015.1092617
Katsoulos P.D., Minas A., Karatzia M.A., Pourliotis K. & Christodoulopoulos G. 2011. Evaluation of a portable glucose meter for use in cattle and sheep. Veterinary Clinical Pathology. 40(2): 245-247. DOI: 10.1111/j.1939-165x.2011.00309.x. DOI: https://doi.org/10.1111/j.1939-165X.2011.00309.x
Koo T.K. & Li M.Y. 2016. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Journal of Chiropractic Medicine. 15(2): 155-163. DOI: 10.1016/j.jcm.2016.02.012. DOI: https://doi.org/10.1016/j.jcm.2016.02.012
Lasson E.D. 2004. Laboratory evaluation of the endocrine pancreas and of glucose metabolism. In: Thrall M.A. (Ed). Veterinary Hematology and Clinical Chemistry. Philadelphia: Lippincott Williams & Wilkins. pp. 431-444.
Mair B., Drillich M., Klein-Jöbstl D., Kanz P., Borchardt S., Meyer L., Schwendenwein I. & Iwersen M. 2016. Glucose concentration in capillary blood of dairy cows obtained by a minimally invasive lancet technique and determined with three different hand-held devices. BMC Veterinary Research. 12: 1-11. DOI: 10.1186/s12917-016-0662-3. DOI: https://doi.org/10.1186/s12917-016-0662-3
Mayer J., Schnellbacher R., Rich E., Divers S.J. & Ward C. 2016. Use of a commercial continuous interstitial glucose monitor in rabbits (Oryctolagus cuniculus). Journal of Exotic Pet Medicine. 25(3): 220-225. DOI: 10.1053/j.jepm.2016.04.006. DOI: https://doi.org/10.1053/j.jepm.2016.04.006
Parkes J.L., Slatin S.L., Pardo S. & Ginsberg B.H. 2000. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 23(8): 1143-1148. DOI: 10.2337/diacare.23.8.1143. DOI: https://doi.org/10.2337/diacare.23.8.1143
Paul-Murphy J. 2007. Critical care of the rabbit. Veterinary Clinics of North America: Exotic Animal Practice. 10(2): 437-461. DOI: 10.1016/j.cvex.2007.03.002. DOI: https://doi.org/10.1016/j.cvex.2007.03.002
Petritz O.A., Antinoff N., Chen S., Kass P.H. & Paul-Murphy J.R. 2013. Evaluation of portable blood glucose meters for measurement of blood glucose concentration in ferrets (Mustela putorius furo). Journal of the American Veterinary Medical Association. 242(3): 350-354. DOI: 10.2460/javma.242.3.350. DOI: https://doi.org/10.2460/javma.242.3.350
Poljičak-Milas N., Kardum-Skelin I., Vudan M., Marenjak T.S., Ballarin-Perharić A. & Milas Z. 2009. Blood cell count analyses and erythrocyte morphometry in New Zealand white rabbits. Veterinarski Arhiv. 79(6): 561-571.
Santos M.A.B.S., Vargas A.M., Rosato P.N., Andrade C.G., Martins C.M. & Petri G. 2022. Evaluation of three human-use glucometers for blood glucose measurement in dogs. Veterinary Medicine International. 24: 1-9. DOI: 10.1155/2022/9112961. DOI: https://doi.org/10.1155/2022/9112961
Seleri P., Di Girolamo N. & Novari G. 2014. Performance of two portable meters and a laboratory analyzer for blood glucose concentration measurement in rabbits. Journal of the American Veterinary Medical Association. 245(1): 87-98. DOI: 10.2460/javma.245.1.87. DOI: https://doi.org/10.2460/javma.245.1.87
Silva K.G., Rotta I., Costa L.B. & Sotomaior C.S. 2020. Comparison of 2 portable human glucometers for the measurement of blood glucose concentration in White New Zealand rabbits. Journal of Veterinary Diagnostic Investigation. 32(5): 683-688. DOI: 10.1177/1040638720946702. DOI: https://doi.org/10.1177/1040638720946702
Tennent-Brown B.S., Koenig A., Williamson L.H. & Boston R.C. 2011. Comparison of three point-of-care blood glucose meters for use in adult and juvenile alpacas. Journal of the American Veterinary Medical Association. 239(3): 380-386. DOI: 10.2460/javma.239.3. DOI: https://doi.org/10.2460/javma.239.3.380
Tonyushkina K. & Nichols J.H. 2009. Glucose meters: a review of technical challenges to obtaining accurate results. Journal of Diabetes Science and Technology. 3(4): 971-980. DOI: 10.1177/193229680900300446. DOI: https://doi.org/10.1177/193229680900300446
Wess G. & Reusch C. 2000. Evaluation of five portable blood glucose meters for use in dogs. Journal of the American Veterinary Medical Association. 216(2): 203-209. DOI: 10.2460/javma.2000.216.203. DOI: https://doi.org/10.2460/javma.2000.216.203
Zar J.H. 2009. Biostatistical analysis. 5th edn. Saddle River: Prentice Hall, pp.943-944.
Zueger T., Schuler V., Stettler C., Diem P. & Christ E.R. 2012. Assessment of three frequently used blood glucose monitoring devices in clinical routine. Swiss Medical Weekly. 142: w13631. DOI: 10.4414/smw.2012.13631. DOI: https://doi.org/10.4414/smw.2012.13631
Additional Files
Published
How to Cite
Issue
Section
License
Copyright (c) 2024 Taíse Fuchs, André Saldanha, Gabriel Aikawa Dusi, Nina da Cunha Medeiros, Bianca Ribeiro de Sales, Eloisa Muehlbauer, Leticia Alves Wrobel, Carolina Zaghi Cavalcante, Cláudia Turra Pimpão

This work is licensed under a Creative Commons Attribution 4.0 International License.
This journal provides open access to all of its content on the principle that making research freely available to the public supports a greater global exchange of knowledge. Such access is associated with increased readership and increased citation of an author's work. For more information on this approach, see the Public Knowledge Project and Directory of Open Access Journals.
We define open access journals as journals that use a funding model that does not charge readers or their institutions for access. From the BOAI definition of "open access" we take the right of users to "read, download, copy, distribute, print, search, or link to the full texts of these articles" as mandatory for a journal to be included in the directory.
La Red y Portal Iberoamericano de Revistas Científicas de Veterinaria de Libre Acceso reúne a las principales publicaciones científicas editadas en España, Portugal, Latino América y otros países del ámbito latino